This machine is based on the new international standards and national standards and design and manufacture of non-metallic material test instrument, mainly used for plastic, hard rubber, nylon, electrical insulating materials, long fiber reinforced composite materials, high strength thermosetting laminating materials and other non-metallic materials of thermal deformation temperature and the measurement of VECA softening point temperature.
Instrument features:
Adopt high precision temperature control meter to display and control temperature, digital display dial meter to display displacement, displacement precision 0.001mm, simple structure, easy to operate.
Compliance with standards:
Standard no | The name of the standard |
GB/T 1633 | Determination of Vicat softening temperature (VST) |
GB/T 1634.1 | Determination of deformation temperature under load for plastics (general test method) |
GB/T 1634.2 | Determination of deformation temperature under load for plastics (plastics, hard rubber and long fiber reinforced composites) |
GB/T 1634.3 | Determination of deformation temperature under load for plastics (high strength thermosetting laminated materials) |
GB/T 8802 | Thermoplastic pipes and fittings - Determination of VEKA softening temperature |
ISO 2507, ISO 75, ISO 306,ASTM D1525,ASTM D648, etc. |
Working principle:
Definition of VEKAT (VST) : the sample is placed in a liquid medium or heating box, and the temperature of the standard pressing needle under the action of (50+1) N force is measured when it is pressed into the sample 1mm cut from the pipe or pipe fittings.
Definition of thermal deformation (HDT) : the standard specimen is subjected to a constant three-point bending load in a horizontal (preferably) or lateral manner to produce one of the bending stresses specified in the relevant section of GB/T 1634. Under the condition of constant temperature rise, the temperature at which the standard deflection corresponding to the specified bending strain increment is measured is reached.
Technical parameters:
XRW-300A thermal deformation, VEKA softening point temperature tester | |
Extraction method of sample holder | Manually extract |
The control mode | LCD temperature control meter control |
Temperature control range | At room temperature to 300 ℃ |
Heating rate | A speed: 5±0.5℃/6min; B speed: 12±1.0℃/6min |
Temperature accuracy | ±0.5℃ |
Temperature measuring point | 1 |
Sample location | 3 |
Deformation resolution | 0.001mm |
Deformation measurement range | 0~10mm |
Sample support span | 64mm、100mm(American standard adjustable size) |
Accuracy of deformation measurement | 0.005mm |
The heating medium | Methyl silicone oil; Flash point above 300℃, below 200 centimes (provided by customers) |
Cooling method | Natural cooling above 150℃, water cooling or natural cooling below 150℃ |
Size of instrument | 700mm×600mm×1400mm |
The required space | 1m front and rear, 0.6m left and right |
The power supply | 4500VA 220VAC 50H |
Instrument configuration:
XRW-300A thermal deformation, VEKA softening point temperature tester | |||
The host | 1 | ||
The power cord | 1 | ||
Sample frame | 3 | ||
Digital display dial meter | 3 | ||
Thermal deformation head | 3 | ||
D conformity of needle | 3 | ||
Assembly tools | 1 | ||
Farmar box | A code | 1000g | 12 |
B code | 900g | 3 | |
C code | 500g | 3 | |
D code | 200g | 6 | |
E code | 100g | 3 | |
F code | 50g | 3 | |
G code | 20g | 6 | |
H code | 10g | 3 | |
I code | 5g | 3 | |
T code | 2g | 6 | |
W code | 60(load bar) | 3 |